Bubble Point vs. Absolute Rating: What a Micron Number Actually Means
The same sintered disc can be sold as 3 μm, 5 μm or 10 μm by three suppliers - and all three can be technically honest. Here is why.
If you compare quotations for the same sintered filter, the rating is often the first number that differs. One supplier says 3 μm, another 5 μm, a third 10 μm, and the prices differ accordingly. The parts may be near-identical. The difference is usually not dishonesty - it is three different definitions being used as if they were one.
Three definitions, one word
| Basis | Definition | Typical use |
|---|---|---|
| Nominal rating | The particle size at which the filter retains a stated percentage (often 90-98%) of a test dust. A statistical figure, not a hard cutoff. | General industrial filtration, where "mostly retained" is acceptable |
| Absolute rating | The largest spherical particle that will pass, usually expressed as the pore size at which no passage occurs. Closer to a hard limit, but still depends on the test method. | Protective filtration, instrument air, hydraulics |
| Bubble point | The pressure at which gas first breaks through a liquid-wetted filter. A physical measurement that can be converted to an equivalent pore diameter. | Verification, incoming inspection, batch consistency |
Why bubble point is the number to ask for
Nominal and absolute ratings are results of a test with particles, and particle tests are slow, messy and sensitive to how the test dust is prepared. Bubble point is a direct physical measurement: wet the filter with a liquid of known surface tension, apply gas pressure from one side, and record the pressure at which the first continuous stream of bubbles appears. That pressure relates to the largest pore through the capillary equation:
ΔP = 4 σ cosθ / d
where σ is the liquid surface tension, θ the contact angle and d the pore diameter. With a wetting liquid that fully wets the surface (cosθ ≈ 1), the conversion is straightforward - which is why bubble point is the standard way to verify a sintered part's rating without a particle challenge.
The catch: bubble point measures the largest pore, not the average
A single oversized pore at the edge of a plate can dominate the bubble point while the bulk of the material is much finer. This is why a bubble point result should always be read together with permeability: a part with an excellent bubble point but abnormally high permeability probably has a local defect; a part with a modest bubble point and low permeability is probably uniformly fine but thicker than necessary.
Why the same disc can be 3, 5 or 10 μm
- Different test liquids. Water, alcohol and proprietary wetting agents have different surface tensions; the same pore gives different bubble point pressures.
- Different conversion conventions. Some suppliers convert using the largest pore, some use a mean pore from a pore-size distribution.
- Different structures. A graded multilayer disc (fine layer on a coarse support) has a fine bubble point but a rating that depends on which face is tested.
- Different definitions. 3 μm absolute and 3 μm nominal are not the same filter, and a 3 μm nominal can easily pass particles larger than 3 μm.
How to compare quotations fairly
- Ask for the rating and its basis: nominal, absolute, or bubble-point-derived.
- Ask for the bubble point value, the test liquid, and whether the test was on the whole part or a coupon.
- Ask for air permeability at a stated pressure drop and area - it is the cheapest way to detect a part that is "fine on paper, coarse in fact".
- For critical duty, ask for a third-party report from an accredited laboratory on the actual batch, not a type-test from years ago.
What this means for your specification
If your process needs a genuine hard limit - protecting a downstream instrument, guaranteeing sterile air - specify absolute rating and a maximum bubble point pore size, and require the value on the batch certificate. If your process only needs bulk removal, a nominal rating with a permeability figure is honest and cheaper. Writing the basis on the drawing costs one line and removes most of the argument later.
Questions we get asked
- Is a smaller micron rating always better?
- No. A finer rating means lower flow for the same area, faster blinding and more frequent cleaning. The correct rating is the coarsest one that protects what is downstream - not the finest one available.
- Can bubble point be converted to an exact pore size?
- It gives an equivalent pore diameter for the largest through-pore, assuming full wetting. It is an excellent comparator between batches and suppliers, but it is not a full pore-size distribution.
- Why does my supplier's 5 μm filter pass particles larger than 5 μm?
- Because the rating is probably nominal, based on a retention percentage rather than a hard cutoff. Ask for the basis and for the bubble point value; if the answer is vague, the number is nominal.
- Do you test every part or a sample?
- Bubble point and permeability are measured per batch on samples from the batch; dimensional inspection is per part against the drawing. For critical orders, every part can be bubble-point tested - state it on the order.
Send both rating figures and we will tell you whether they describe the same filter - no obligation, and we will say so if the other quote is the better fit. [email protected]